Black 8-liter Class B autoclave with USB
Warranty details.
Safety
- The cycle does not start if the door is not properly closed.
- The automated safety system with a double hydraulic locking device prevents the door from opening while the chamber is under pressure.
- Safety valves prevent pressure buildup in the pressure generator and the camera.
- A safety thermostat prevents overheating in the chamber and the pressure generator.
- An automatic safety lock prevents the door from accidentally opening during the sterilization process, and the process protector automatically cuts off the power supply if there is a short circuit or if the power supply is unstable.
- A failure alarm indicates a failed cycle and provides a detailed error code.
- A door alarm indicates that the door is not properly secured.
- A low-water alarm indicates when the water tank is empty or overfilled.
Technical specifications
- Voltage: 220-240 VAC; 50/60 Hz; 10A
- Chamber dimensions: 170 x 320 mm
- Tray size: 130 x 265 mm
- Circuit breaker: F16A / 400 V
- 2000 W
- Rated power: 1500 VA
- Sterilization temperature: 121 °C / 134 °C
- Distilled water tank capacity: 2.5 L
- Operating temperature: 5 °C / 40
- Relative humidity: Max. 80%, non-condensing
- External dimensions: 420 x 370 x 525 mm
- Net weight: 37 kg
- Maximum noise level: <70 dB
- Atmospheric pressure: 76 kPa / 106 kPa
Maintenance and error message guide for your sterilizer
Icanclave autoclave: maintenance and error message guide
Components
- Distilled water tank
- LCD screen
- Control panel
- Main switch
- Key for draining distilled and used water
- USB port
- Safety valve
- Ventilation grille
- Locking system
- Power cable
- Label
Included accessories
- Sterilizer
- USB
- 2 instrument trays
- Tray support
- Handle for removing trays
- Door adjustment tool
- 2 drain hoses
- Door cover
- Instruction manual
Regulations
- 93/42 / EEC Medical devices directive.
- 97/23 / EEC Pressure equipment directive.
- EN 13060 Small steam sterilizers.
- EN 61326, EN 61000 EMC.
- EN 61010-1, UL 61010-1 safety requirements.
- EN 61010-2-040, IEC 61010-2-040 Particular requirements for steam sterilizers used for processing medical materials.
Autoclave process step by step
1. Open the door, take out the tools and accessories located in the sterilization chamber, remove the packaging and clean them.
2. Connect the power cord to a suitable power source.
3. Connect the printer (sold separately).
4. Switching on: the main switch is located on the front of the device, at the bottom right. Once switched on, the LCD screen will light up and the basic autoclave operating symbols, such as door position, water level, sterilization program, date, and time, will appear.
5. Remove the top cover and then fill the distilled water tank. If an audible signal sounds, it means you have exceeded the permitted water level in the tank. Empty the excess water until the alarm stops.
6. Preparing the material to be sterilized: place the instruments on the various trays and separate them according to their materials. If the instruments are not made of stainless steel, use sterilization pouches. These must not come into contact with each other. Check that all instruments are properly placed so that water does not remain in hollow parts. Do not overload the instrument trays or place them in direct contact with the walls of the sterilization chamber. Do not use paper clips or needles to avoid damaging or causing a malfunction of the autoclave.
7. Select the sterilization program.
8. Start the sterilization cycle.
After selecting the sterilization program, place the instruments on the trays in the chamber using the provided plastic tongs.
Once the instruments are properly placed inside, close the door. Caution: make sure the door is properly closed, otherwise the autoclave will not be able to operate.
9. Start the sterilization program.
Press the start button, and the autoclave will begin to operate for approximately 30-75 minutes.
10. End of the sterilization cycle.
Once the sterilization cycle is complete, the printer (not included) will automatically print the full cycle report.
Once the pressure has returned to 0, you can open the door and remove the instruments. The autoclave will alert you with an audible signal once the sterilization cycle is complete.
Caution: Use the plastic tongs to remove the instruments.
IMPORTANT: Do not reuse the wastewater after emptying the tank.
It is important to keep the packaging box in case of after-sales service